ClearanceBrief
Recall Z-1627-2026 ยท ITX Transducer, Ultrasonic, Diagnostic

Philips Ultrasound, Inc

Philips X3-1 Ultrasound Transducer

DateMar 24, 2026 (initiated Sep 5, 2025)
StatusOpen, Classified
ReasonTo provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Root causeDevice Design
ActionAn URGENT Medical Device Correction Notice, dated 8/19/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; Th
Quantity17 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042540PHILIPS IU22 ULTRASOUND SYSTEM
Philips Ultrasound, Inc., Oct 4, 2004