Recall Z-1637-2026 ยท NHJ Device, Positive Pressure Breathing, Intermittent

Baxter Healthcare Corporation

Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system.

DateApr 1, 2026 (initiated Feb 18, 2026)
StatusOpen, Classified
ReasonUrgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.
Root causeProcess control
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 2/18/2026 via email. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Continue your Volara therapy as prescribed by your doctor. 2. Refer to the enclosed IFU addendum for additional information related to use in the home care environment. The following warning was added for consideration related to this issue: Warning - Patients that may have difficulty clearing secretions from the upper airway may require specialized therapy (for example, separ
Quantity19 units
DistributionFlorida
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192143Maximus System
Hill-Rom Services Pte, Ltd., Feb 14, 2020
K200988Maximus System
Hill-Rom Services Pte, Ltd., May 26, 2020