ClearanceBrief
Recall Z-1583-2026 ยท FON Bag, Drainage, With Adhesive, Ostomy

ConvaTec, Inc

Convatec, EsteemBody Drainable Pouch REF:423653 10-35mm

DateMar 19, 2026 (initiated Feb 16, 2026)
StatusOpen, Classified
ReasonDrainable large pouch may leak due to manufacturing issue.
Root causeProcess control
ActionOn February XX, 2026 Convatec issued a updated Urgent Medical Device Recall Notification via E-Mail. The initial notification was issued February 16, 2026. Convatec asked consignees to take the following actions: 1. Review your personal supplies and determine if you are affected by this communication. 2. Discontinue use and dispose of the affected product. 3. Complete Response Form and return to Convatec within 30 days. 4. Acknowledgment of this action is critical. If you re not able to provide a written response liaise with your distributor or Convatec Customer Services can support. 5.
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Canada, China, Czech Republic, Denmark, Finland, Germany, Ireland, Italy, Japan, Netherlands, Norway, Poland, Slovakia, Sweden, Switzerland, UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.