ClearanceBrief
Recall Z-1643-2026 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Olympus Corporation of the Americas

Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

DateMar 25, 2026 (initiated Feb 25, 2026)
StatusOpen, Classified
ReasonOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Root causeNonconforming Material/Component
ActionOn February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this letter through the Olymps portal 5. Please forward this notice to other users who may have the affected products if you have further distributed it.
Quantity7 units
DistributionWorldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023492EVEREST BIPOLAR CUTTING FORCEPS AND GYRUS BIPOLAR CUTTING FORCEPS
Gyrus Medical, Inc., Nov 13, 2002