ClearanceBrief
Recall Z-1675-2026 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;

DateMar 27, 2026 (initiated Mar 3, 2026)
StatusOpen, Classified
ReasonPhilips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Root causeDevice Design
ActionOn March 3, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue and follow the instructions below. 2. Perform the daily check before starting the first procedure, as described in the device labelling 3. Should you experience a foot switch activation issue, please report it to your local Philips representative. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device
Quantity670 units (106 US, 564 OUS)
DistributionDomestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.