Recall Z-1564-2026 ยท PUI Drape, Surgical, Exempt

Medline Industries, LP

Medline Surgical Drapes: MDTBTCS6070GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTBTCS6090GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTDXUNDBGN DBD-DRAPE,UNDERBUTTOCK,28 X 50.5,XALT GR MDTMAYOXGN DBD-DRAPE,COVER,MAYO STAND,RESISTAT/XALT MDTZ1003656GN DBD-POCKET SHEET,RESISTAT/XALT

DateMar 12, 2026 (initiated Jan 28, 2026)
StatusOpen, Classified
ReasonFabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
Root causeLabeling Change Control
ActionOn January 28, 2026, the firm notified customers of the issue through Urgent Medical Device Recall letters. Users were provided with updated laundering instructions and general inspection guidelines.
Quantity23,568
DistributionDomestic distribution nationwide US. International distribution to Panama, United Arab Emirates, Virgin Islands, Saudi Arabia, India, Chile, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.