Recall Z-1567-2026 ยท GEH Unit, Cryosurgical, Accessories

Erbe USA Inc

Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use

DateMar 20, 2026 (initiated Feb 12, 2026)
StatusOpen, Classified
ReasonProbes may rupture/burst during activation
Root causeProcess control
ActionOn May 5, 2026, ERBE USA issued a "Urgent Medical Device Recall Expansion notification via E-Mail. The expansion was issued due to new lot numbers that may cause the same failure mode. The original notification was issued on Feb. 12, 2026. ERBE asked consignees to take the following actions: 1. Discontinue use of the products with the affected lot numbers. 2. Examine your inventory, remove, and quarantine any identified devices with the affected lot numbers 3. Ensure all applicable personnel within the vicinity of device activation carefully read the content of this notification. If affected
Quantity8,284 units
DistributionUS Nationwide distribution, including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190651ERBECRYO 2 Cryosurgical Unit; ERBECRYO 2 1-pedal footswitch, Extension cable f. ERBECRYO 2 footswitch; Flexible Cryoprob
Erbe Elektromedizin GmbH, Jan 7, 2020