Recall Z-2161-2026 ยท DRM Compressor, Cardiac, External

Jolife AB

LUCAS 2 Model Number: 99576 Catalog Numbers: 99576-000024 and 99576-000011 UDI-DI code: 00883873904565 LUCAS 3 Model Number: 99576 Catalog Number: 99576-000043 UDI-DI code: 00883873861875 LUCAS 3.1 Model Number: 99576 Catalog Number: 99576-000063 UDI-DI code: 00883873834329 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acut

DateMay 11, 2026 (initiated Apr 13, 2026)
StatusOpen, Classified
ReasonDue to a subset of chest compression systems/devices where the service performed on the systems were not appropriately documented (The documents for the service performed to justify the passing results were not include in the service records).
Root causeUnder Investigation by firm
ActionOn 04/13/2026, the firm sent via US Postal an "Urgent: Medical device correction" letter informing customers that through complaint reviews, Stryker discovered a subset of work orders performed between July 2023 through November 2025 with Performance Inspection Procedures (PIP) recorded as PASS without the corresponding objective evidence properly documented in the work orders. Customers are instructed to: 1. Immediately check internal inventory to locate the affected product as listed. 2. Return the business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification.
Quantity153 systems
DistributionU.S. Nationwide distribution in the states of AL, AZ, CA, FL, HI, IA, KS, KY, LA, MA, MD, MN, NC, NV, OH, SC, TX, UT, VA, WA, and WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090422LUCAS 2
Jolife AB, Jun 17, 2009
K161768LUCAS 3 Chest Compression System
Jolife AB, Nov 9, 2016