Recall Z-1591-2026 ยท OZD Temporary Non-Roller Type Left Heart Support Blood Pump

Abiomed, Inc.

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packa

DateMar 27, 2026 (initiated Feb 16, 2026)
StatusOpen, Classified
ReasonPotential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Root causeSoftware design
ActionAbiomed notified international consignees on about 02/16/2026 via emailed CRITICAL PRODUCT CORRECTION letter. Consignees were instructed to cease use of the Automated Impella Controller (AIC) with an Impella 5.5 S1 (product code 0550-0007) unless AIC software is updated to v10.1 or above, identify current software in the lower left corner of the initial AIC display screen, contact your local Abiomed representative to schedule your software update, review and complete the customer response form provided and post this notification in a visible area for awareness. Consignees were also instructed
Quantity62 units
DistributionInternational distribution in the country of Australia, Canada, France, Germany, Kuwait.
FDA recordOpen on fda.gov