Recall Z-1610-2026 ยท ODC Endoscope Channel Accessory

Olympus Corporation of the Americas

Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.

DateMar 20, 2026 (initiated Feb 12, 2026)
StatusOpen, Classified
ReasonThe MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Root causeReprocessing Controls
ActionOlympus notified consignees on about 02/12/2026 via URGENT: MEDICAL DEVICE CORRECTION letter. Consignees were instructed to immediately cease reprocessing the MAJ-1443 and MAJ-1444 valves in an OER-Pro or OER-Elite and ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the full extent of the MAJ-1443 and MAJ-1444 reprocessing steps as detailed in the provided Instructions for Use (IFU). The updated IFU can also be accessed electronically through OlympusConnect.com. Additionally, consignees were instructed to replace any existing
Quantity47,383 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K250949Suction Valve (MAJ-1443); Air/Water Valve (MAJ-1444)
Olympus Medical Systems Corporation, Jul 16, 2025