Recall Z-1484-2026 · JIR Indicator Method, Protein Or Albumin (Urinary, Non-Quant.)

Siemens Healthcare Diagnostics, Inc.

Atellica CH Urine Albumin (UAlb). Material Number: 11537225

DateMar 2, 2026 (initiated Jan 20, 2026)
StatusOpen, Classified
ReasonFalsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Root causeUnder Investigation by firm
ActionSiemens Healthineers notified consignees on about 01/20/2026 via mailed FedEx letter titled Urgent Medical Device Correction. Consignees were informed that as temporary mitigation, for any sample with a UAlb result above 5.0 mg/dL (50 mg/L), reanalyze the sample with a x10 dilution. If the diluted result is consistent with the initial result, either result may be reported. If the diluted result is inconsistent with the initial result, the diluted result should be reported. If you prefer to automate this dilution, please refer to your internal laboratory procedures. Consignees were instructed
Quantity4,885 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.