Recall Z-1475-2026 ยท LLZ System, Image Processing, Radiological

GE Healthcare

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; System, Image Processing, Radiological

DateFeb 27, 2026 (initiated Jan 16, 2026)
StatusOpen, Classified
ReasonUnder certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,
Root causeDevice Design
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 01/16/2026 by letter delivered via traceable means. The notice explained the issue and risk, and requested the following: Actions to be taken by Customer/User: "To determine whether you have ZFP Patient Timeline configured at your site, please contact your site administrator. " If you do not have the ZFP Patient Timeline, you can continue to use Centricity Universal Viewer Zero Footprint Client (ZFP) DICOM viewer." "Review all patient studies that were reviewed using the ZFP Patient Timeline to ensure t
Quantity250 units
DistributionWorldwide distribution - US Nationwide and the countries of Belarus, Brazil, Canada, China, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Lebanon, Malaysia, Netherlands, Philippines, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spai
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131977CENTRICITY UNIVERSAL VIEWER ZERO FOOTPRINT CLIENT
GE Healthcare, Sep 17, 2013