Recall Z-1549-2026 ยท HWC Screw, Fixation, Bone

Medartis AG

Brand Name: APTUS Product Name: 2.5 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5750.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. Component: Not applicable.

DateMar 10, 2026 (initiated Jan 9, 2026)
StatusOpen, Classified
ReasonMix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
Root causeLabeling mix-ups
ActionFirm began notifying affected customers on January 13, 2026 via email. Customers were instructed to return impacted products to Medartis Inc.
Quantity0
DistributionWorldwide - US Nationwide distribution in the states of Indiana and Pennsylvania and the countries of Austria, Belgium, Switzerland, Czech Republic, Germany, Spain, Finland, France, Israel, Italy, Japan, and Poland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051567APTUS TITANIUM OSTEOSYNTHESIS SYSTEM
Medartis, Inc., Aug 23, 2005