ClearanceBrief
Recall Z-1606-2026 ยท HWC Screw, Fixation, Bone

I.T.S. GmbH

HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 3

DateMar 20, 2026 (initiated Jan 23, 2026)
StatusOpen, Classified
ReasonUpdated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Root causeNo Marketing Application
ActionITS initially notified its importer/distributor on about 01/23/2026 via email. Instructions were to cease distribution, and further information will be provided. An additional notification was sent to distributors and consignees on about 02/04/2026 via email. Instructions included to ensure any IFU different to the provided one is removed from systems and/or destroyed, notify clinical staff managing patients with implants, contact I.T.S. GmbH for any questions or to discuss proper scanning conditions for implanted devices in the interim while they work on labeling and IFU updates for the devi
Quantity623 units
DistributionUS Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131722I.T.S. EXTREMITY FIXATION SYSTEMS
I.T.S. GmbH, Aug 7, 2013