ClearanceBrief
Recall Z-1423-2026 ยท PUI Drape, Surgical, Exempt

Medline Industries, LP

Medline Surgical Drapes: 1) SHEET,DRAPE,40X58,STERILE, Model Number: DYNJP2410; 2) DRAPE,SHEET,ECONOMY,40X58",ST,20/CS, Model Number: DYNJP2411; 3) SHEET,DRAPE, 40X70,STERILE, Model Number: DYNJP2412; 4) DRAPE,MEDIUM,TAPE,40X70",ST,50/CS, Model Number: DYNJP2412T; 5) SHEET,DRAPE,53X77,STERILE, Model Number: DYNJP2414; 6) DRAPE,SHEET,3/4,53X77",ST,20/CS, Model Number: DYNJP2414R; 7) DRAPE,SHE

DateFeb 17, 2026 (initiated Jan 7, 2026)
StatusOpen, Classified
ReasonMedline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Proxima Sterile Surgical Gowns, Packs, and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected prod
Quantity14,927,765 units
DistributionWorldwide distribution - US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.