ClearanceBrief
Recall Z-1349-2026 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X

DateFeb 11, 2026 (initiated Jan 26, 2026)
StatusOpen, Classified
ReasonUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Root causeDevice Design
ActionOn January 26, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. " Read and follow the additional instructions in Appendix A regarding the use and configuration of the Reset Geometry function. " Keep Appendix A together with the documentation of the system. Ensure that the letter is in a place likely to be seen/viewed. " Confirm the Reset Geometry function that best suits your clinical workflow. If required, you may request Philips
Quantity5136 units (1743 US, 3393 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017
K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018
K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020