Recall Z-1346-2026 ยท OWB Interventional Fluoroscopic X-Ray System
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.X
| Date | Feb 11, 2026 (initiated Jan 26, 2026) |
| Status | Open, Classified |
| Reason | Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active. |
| Root cause | Device Design |
| Action | On January 26, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. " Read and follow the additional instructions in Appendix A regarding the use and configuration of the Reset Geometry function. " Keep Appendix A together with the documentation of the system. Ensure that the letter is in a place likely to be seen/viewed. " Confirm the Reset Geometry function that best suits your clinical workflow. If required, you may request Philips |
| Quantity | 655 units (106 US, 549 OUS) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, |
| FDA record | Open on fda.gov |