Recall Z-1826-2026 ยท ITI Wheelchair, Powered
WHILL, INC.
WHILL Model C2, with software versions before HMI: 2.22 / MC: 1.13; device is an indoor/outdoor battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair.
| Date | Apr 24, 2026 (initiated Dec 29, 2025) |
| Status | Open, Classified |
| Reason | Cyber vulnerability of BLE communication was found by CISA VDP program. |
| Root cause | Device Design |
| Action | Whill, Inc. issued a Medical Device Recall notice to its sole consignee on 3/23/2026 via email. The notice explained the issue, risks to health, and how to recognize the failure of the device. The notice also provided the following instructions for the distributor: Actions to be taken by the Distributor: 1. Cease all distribution of the devices until the corrective firmware update (2.24 for Model C2 and 2.25 for Model F) has been successfully installed. 2. Notify all customers with previous versions of Model C2 and Model F firmware of this issue. You may use this notice with the User Respo |
| Quantity | 4685 units |
| Distribution | US: Kansas |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K213383 | WHILL Model C2 Whill, Inc., Nov 9, 2021 |