Recall Z-1496-2026 ยท FMF Syringe, Piston

Becton Dickinson & Company

10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.

DateMar 6, 2026 (initiated Jan 27, 2026)
StatusOpen, Classified
ReasonDuring the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
Root causeUnder Investigation by firm
ActionBecton Dickinson notified consignees on about 01/27/2026 via emailed URGENT: Medical Device Recall (Correction) letter. Consignees were instructed to visually verify that the correct syringe is selected for the intended clinical application prior to use and that it forms a secure connection, destroy any luer slip syringes found in inventory, circulate the notification to all personnel who need to be aware or to any organization where the potentially affected products have been transferred, and complete and return the provided Customer Response Form. Distributors and kit manufactures were ins
Quantity204,000 units
DistributionUS Nationwide distribution in the states of California and Texas.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980987BECTON DICKINSON SYRINGE