ClearanceBrief
Recall Z-1342-2026 ยท OLO Orthopedic Stereotaxic Instrument

Alphatec Spine, Inc.

ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 266-01-000-10-N UDI-DI code: 00190376476179; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker.

DateFeb 11, 2026 (initiated Jan 2, 2026)
StatusOpen, Classified
ReasonDue a design issue where the navigated array connection geometry is incorrect.
Root causeDevice Design
ActionOn 01/02/2026, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customers informing them that Alphatec Spine has been notified of complaints regarding LIF Navigated Osteotome instruments appearing to have incorrect orientation when used with the Medtronic StealthStation". The instrument is correctly identified and displays the correct navigated length; however, the projected image on the Medtronic StealthStation" is rotated by 90. Customers are instructed to: "Review their inventory to determine if any of these affected devices are in their possession. If a device is
Quantity7 units
DistributionU.S. Nationwide distribution in the states of AZ, CA, FL, IL, IN, MA, MI, MO, NC, NH, NY, TN, TX, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223765ATEC Lateral Navigation Instruments
Alphatec Spine, Inc., Mar 23, 2023