ClearanceBrief
Recall Z-1368-2026 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

DateFeb 13, 2026 (initiated Jan 13, 2026)
StatusOpen, Classified
ReasonPumps were released without full testing being performed, including occlusion alarm testing.
Root causeEmployee error
ActionFirm notified the single customer on 1/13/26 via phone call. Customer was asked to work with on-site service technician to coordinate return of affected product.
Quantity5
DistributionStates: Ohio Countries: United States
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K251636Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)
Baxter Healthcare Corporation, Jul 28, 2025