ClearanceBrief
Recall Z-1437-2026 ยท GCP Endoscope, Ac-Powered And Accessories

Olympus Corporation of the Americas

Olympus Cystoscope Outer Sheath; Model/Catalog number: WA22810A;

DateFeb 20, 2026 (initiated Sep 13, 2024)
StatusOpen, Classified
ReasonProduct being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device.
Root causeUnder Investigation by firm
ActionOn January 15, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected products in your inventory. 3. Olympus requests that you acknowledge receipt of this letter through the recall web portal provided. 4. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your WA2281
Quantity633 units
DistributionDistribution US nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K790071ENDOSCOPE AND ACCESSORIES
Olympus Corp., Apr 2, 1979