Recall Z-1427-2026 ยท NEU Marker, Radiographic, Implantable

Stryker Corporation

MOLLI 2 System System includes: MM1000 (Pack of 10) - MOLLI Introducer (8cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195028 MS-00009 (Pack of 10) - MOLLI Introducer (12cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195134 MP2001 - MOLLI 2 Wand UDI (GTIN): 00850024195141 MP2015A - MOLLI OncoPen UDI (GTIN): 00850024195219 MT2001 - MOLLI 2 Tablet UDI (GTIN): 00850024195158 The MOLLI Marker is i

DateFeb 17, 2026 (initiated Jan 21, 2026)
StatusOpen, Classified
ReasonDue to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.
Root causeUnder Investigation by firm
ActionOn 01/21/2026, the firm sent via USPS certified mail an "URGENT: Medical Device Recall" Letter to customers informing them that Stryker has updated the MOLLI 2 System Instructions for Use (IFU) following two Medical Device Reports (MDRs) submitted to the FDA. These reports involved patient injuries associated with MOLLI marker dislodgement or dislocation due to magnetic attraction from magnetized surgical tools. The updated IFU now includes an additional warning that using magnetized surgical tools in close proximity to the marker during localization may lead to marker dislodgement. Customers
Quantity5,864
DistributionWorldwide - U.S. Nationwide distribution including in the states of AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, HI ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY. The countries of Canada, Singapore, Panam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231579MOLLI 2
Molli Surgical, Inc., Oct 26, 2023
K253888MOLLI 2 System
Stryker Endoscopy, Dec 31, 2025