ClearanceBrief
Recall Z-1441-2026 ยท FJL Resectoscope

Olympus Corporation of the Americas

Olympus Resection Sheath, 24 Fr; Model No. A22041A. Resection sheath for urologic applications.

DateFeb 20, 2026 (initiated Nov 11, 2025)
StatusOpen, Classified
ReasonComplaints of the ceramic tip of the resection sheath breaking have been received.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE CORRECTIVE ACTION notice was sent to consignees 1/6/2026. Consignees are instructed to ensure all personnel are familiar with the contents of the notification. Users are to adhere to the Inspection and Testing and Caution sections of the IFU for both the resection/inner sheath, as well as the Inspection Before Each Use section in the Resectoscope System Manual. Additional instructions include having an extra resection sheath and inspect during and after procedure for device fragments. Consignees are to retain a copy of the notification with the IFU. Consignees with any
Quantity3,484 units
DistributionUS Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K931995RESECTOSCOPE SHEATH
Olympus Corp., Jul 23, 1993