Recall Z-1320-2026 ยท NLH Catheter, Recording, Electrode, Reprocessed
Medline Industries, LP
Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diag
| Date | Feb 6, 2026 (initiated Dec 22, 2025) |
| Status | Open, Classified |
| Reason | These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025 |
| Root cause | Under Investigation by firm |
| Action | On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have |
| Quantity | 311 |
| Distribution | Distribution US nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K242225 | Medline ReNewal Reprocessed Abbott Inquiry Steerable Diagnostic Catheter Surgical Instrument Service and Savings, Inc., Nov 29, 2024 |