ClearanceBrief
Recall Z-1470-2026 ยท PYX Temporary Non-Roller Type Right Heart Support Blood Pump

Abiomed, Inc.

Impella RP Flex with SmartAssist. Product Code: 1000323.

DateFeb 27, 2026 (initiated Jan 27, 2026)
StatusOpen, Classified
ReasonDifferential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
Root causeUnder Investigation by firm
ActionAbiomed notified consignees on about 01/27/2026 via UPS delivery. Additionally, the Customer Letter was posted on Heartrecovery.com: https://www.heartrecovery.com/en-us/resources/downloads/field-safety-notification-impella-rp-differential-pressure-sensor. The notifications informed consignees were instructed clinicians that affected units can continue use and they should: 1. Continue to monitor patient hemodynamics with approved diagnostic devices and continue to verify Impella positioning with imaging before clinical interventions. 2. Refer to the P-level flow rates listed in the Instructi
Quantity4,496 units
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Canada, Croatia, Czechia, France, Germany, Hong Kong, Italy, Poland, Serbia, Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.