ClearanceBrief
Recall Z-1270-2026 ยท KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Olympus Corporation of the Americas

Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-431Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

DateFeb 5, 2026 (initiated Jan 7, 2026)
StatusOpen, Classified
ReasonDevices which did not undergo thermoforming could deform and lose performance.
Root causeProcess control
ActionOn or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this lett
Quantity7689 units (7501 US, 188 OUS)
DistributionWorldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and International: TBD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K250945Single Use Preloaded Sphincterotome V (Distal Wire Guided) (KD-VC600 Series); Single Use Sphincterotome V (Distal Wire G
Olympus Medical Systems Corp., Oct 17, 2025