ClearanceBrief
Recall Z-1317-2026 ยท JAK System, X-Ray, Tomography, Computed

Philips North America Llc

Spectral CT; Product Code (REF): 728333;

DateFeb 6, 2026 (initiated Jan 7, 2026)
StatusOpen, Classified
ReasonAs part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Root causeProcess change control
ActionOn or about January 7, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. As a precaution, monitor the gantry for any unusual sounds or vibrations during operation. If unusual sounds are heard or vibrations are observed, immediately call your service contact. In the absence of unusual sounds or vibrations, you may continue to use your system(s) according to the intended use. 2. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until Phil
Quantity283 units
DistributionWorldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egy
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K244008Spectral CT
Philips Medical Systems Technologies , Ltd., May 16, 2025