ClearanceBrief
Recall Z-1354-2026 ยท HIF Insufflator, Laparoscopic

Olympus Corporation of the Americas

Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdominal cavity.

DateFeb 18, 2026 (initiated Jan 16, 2026)
StatusOpen, Classified
ReasonIssue with software algorithm which may lead to overpressure events.
Root causeSoftware design
ActionOlympus notified consignees on about 01/16/2026 via "URGENT: MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to cease usage of the affected product immediately. They are advised to use available alternatives including an existing UHI-4 if available or switching to an insufflation device from another manufacturer. If no alternative is available on site, contact Olympus to arrange a loaner or purchase a replacement. Additionally, consignees were instructed to examine inventory and quarantine any identified devices, immediately cease usage of any UHI, UHI-2, and/or UHI-3 in inventory,
Quantity485 units
DistributionNationwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K014166UHI-3 HIGH FLOW INSUFFLATION UNIT
Olympus Optical Co., Ltd., Jun 25, 2002