Recall Z-1162-2026 ยท KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Boston Scientific Corporation

*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;

DateFeb 5, 2026 (initiated Dec 19, 2025)
StatusOpen, Classified
ReasonIncreased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
Root causeProcess control
ActionOn December 19, 2025, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: - Immediately stop further distribution or use of any remaining affected devices. Remove them from your facility s inventory and segregate them in a secure location until they can be returned to Boston Scientific. " Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. " Share this notice with any health care professional within your organizati
Quantity150 units
DistributionWorldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K233318AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery Syst
Boston Scientific Corporation, Oct 27, 2023