Recall Z-1155-2026 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Edan Diagnostics

Brand Name: EDAN Product Name: Telemetry Transmitter Model/Catalog Number: iT20 Product Description: iT20 collects physiological parameters by ECG cables and SpO2 sensors, then achieves data analyzing and processing. After that, data will be sent to MFM-CMS via Wi-Fi. The parameters supported are ECG, SpO2 and PR. Component: No

DateJan 22, 2026 (initiated Oct 24, 2025)
StatusOpen, Classified
ReasonAs a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Root causeDevice Design
ActionOn 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructe
Quantity1457
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161056Telemetry Transmitter
Edan Instruments, Inc., Feb 3, 2017