Recall Z-1151-2026 ยท HGM System, Monitoring, Perinatal
Edan Diagnostics
Brand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CNS, MFM-CNS Lite Product Description: The MFM-CNS and MFM-CNS Lite are clinical data managing software applications. Both applications manage clinical data of fetal monitoring and uterine activity, and the MFM-CNS additional monitors maternal vital signs. Data are automatically acquired from bedside monitors, for t
| Date | Jan 22, 2026 (initiated Oct 24, 2025) |
| Status | Open, Classified |
| Reason | As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device. |
| Root cause | Device Design |
| Action | On 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructe |
| Quantity | 395 |
| Distribution | Worldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K171178 | Central Monitoring System Edan Instruments, Inc., Sep 6, 2017 |