Recall Z-1147-2026 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Edan Diagnostics

Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM20 Product Description: The iM20 Patient Monitor System (hereinafter called iM20) can perform long-time continuous monitoring of multiple physiological parameters, including ECG, respiration (RESP), non-invasive blood pressure (NIBP), oxygen saturation of the blood (SpO2), pulse rate (PR), temperature (TEMP), invasive pressure

DateJan 22, 2026 (initiated Oct 24, 2025)
StatusOpen, Classified
ReasonAs a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Root causeDevice Design
ActionOn 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructe
Quantity377
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152552Patient Monitor
Edan Instruments, Inc., Apr 29, 2016