Recall Z-1146-2026 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Edan Diagnostics

Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: elite V5, elite V6, elite V8 Product Description: V series Patient Monitor including elite V5, elite V6 and elite V8 which can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abn

DateJan 22, 2026 (initiated Oct 24, 2025)
StatusOpen, Classified
ReasonAs a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Root causeDevice Design
ActionOn 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructe
Quantity342
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160981Patient Monitor
Edan Instruments, Inc., Dec 22, 2016