Recall Z-1144-2026 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Edan Diagnostics
Brand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: iM3s, iM3As, iM3Bs, iHM3s Product Description: The iM3s series vital signs monitors including iM3s/iM3As/iM3Bs/iHM3s are intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters of adults and pediatrics. Component: No
| Date | Jan 22, 2026 (initiated Oct 24, 2025) |
| Status | Open, Classified |
| Reason | As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device. |
| Root cause | Device Design |
| Action | On 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructe |
| Quantity | 136 |
| Distribution | Worldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico. |
| FDA record | Open on fda.gov |