Recall Z-1142-2026 ยท DQA Oximeter

Edan Diagnostics

Brand Name: EDAN Product Name: Vital Signs Monitor Model/Catalog Number: M3 Product Description: The Vital Signs Monitor is a portable device intended for use by health care professionals. The monitor could provide the monitoring of physiological parameters such as the non-invasive blood pressure (NIBP), the oxygen saturation of the blood (SpO2) and Rectal/Oral/Auxiliary/Ear temperature measuremen

DateJan 22, 2026 (initiated Oct 24, 2025)
StatusOpen, Classified
ReasonAs a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Root causeDevice Design
ActionOn 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructe
Quantity13934
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131818VITAL SIGNS MONITOR
Edan Instruments, Inc., Aug 30, 2013