Recall Z-1306-2026 · SET Immunoassay Blood Test For Amyloid Pathology Assessment

Fujirebio Diagnostics, Inc.

Product Name: Lumipulse G -Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1-42-N Plasma Calibrators: The -Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported

DateFeb 5, 2026 (initiated Dec 11, 2025)
StatusOpen, Classified
ReasonCustomers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Root causeUnder Investigation by firm
ActionOn or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain th
Quantity128 units
DistributionUS Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K242706Lumipulse G pTau217/ß-Amyloid 1-42 Plasma Ratio
Fujirebio Diagnostics,Inc., May 16, 2025