Recall Z-1191-2026 ยท FDE Laparoscopy Kit

Medline Industries, LP

Medline medical procedure convenience kits labeled as: 1) SMJ LAP CHOLE PACK-LF, Kit SKU DYNJ19974P; 2) DEPAUL ABDOMINAL ENDOSCOPY-LF, Kit SKU DYNJ20027K; 3) LAPAROSCOPY PACK-LF, Kit SKU DYNJ20468J; 4) SMJ LAPAROSCOPY PACK, Kit SKU DYNJ25763N; 5) SMJ LAPAROSCOPY PACK, Kit SKU DYNJ25763O; 6) BWNBORO LAP CHOLE PACK-LF, Kit SKU DYNJ32836B; 7) MINOR SET UP PACK, Kit SKU DYNJ43272K; 8) PK, GEN

DateJan 28, 2026 (initiated Dec 12, 2025)
StatusOpen, Classified
ReasonMedline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root causeNonconforming Material/Component
ActionMedline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected produ
Quantity456 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.