Recall Z-1190-2026 ยท OGR Ear, Nose, And Throat Surgical Tray
Medline Industries, LP
Medline medical procedure convenience kits labeled as: 1) MIDDLE EAR CDS, Kit SKU CDS984347O; 2) HEAD AND NECK CDS-LF, Kit SKU CDS984857N; 3) NECK PACK-LF, Kit SKU DYNJ0966130K; 4) NECKLINE PACK-LF, Kit SKU DYNJ51440B; 5) THYROID PACK, Kit SKU DYNJ58742J; 6) ENT I, Kit SKU DYNJ63781B; 7) NECK PACK, Kit SKU DYNJ82751B; 8) NOSE PACK, Kit SKU DYNJ82753C; 9) WMC COCHLEAR IMPLANT ADD ON, Kit
| Date | Jan 28, 2026 (initiated Dec 12, 2025) |
| Status | Open, Classified |
| Reason | Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive. |
| Root cause | Nonconforming Material/Component |
| Action | Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected produ |
| Quantity | 40 units |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.