Recall Z-1189-2026 ยท OFG Ct Biopsy Tray

Medline Industries, LP

Medline medical procedure convenience kits labeled as: 1) BREAST-HERNIA-PORT CDS-LF, Kit SKU CDS984853L; 2) BREAST BIOPSY PACK, Kit SKU DYNJ45641B; 3) DERMATOLOGY PACK, Kit SKU DYNJT3658A.

DateJan 28, 2026 (initiated Dec 12, 2025)
StatusOpen, Classified
ReasonMedline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root causeNonconforming Material/Component
ActionMedline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected produ
Quantity1,928 total
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.