ClearanceBrief
Recall Z-1134-2026 ยท EOQ Bronchoscope (Flexible Or Rigid)

Intuitive Surgical, Inc.

Brand Name: Ion Endoluminal System Product Name: Ion Vision Probe Bag Model/Catalog Number: 490127 Software Version: N/A Product Description: The Ion Endoluminal System is a robotically assisted platform for minimally invasive peripheral lung biopsy. Ion consists of a System Cart and Controller that are used to navigate a flexible Catheter within the pulmonary tract using endoscopic visualization

DateJan 21, 2026 (initiated Dec 4, 2025)
StatusOpen, Classified
ReasonDue to probe bags that may have a compromised or incomplete sterile pouch seal.
Root causeProcess control
ActionOn 12/19/2025, the firm emailed to customer an "Urgent Medical Device Recall" letter informing them that Intuitive has identified that some Vision Probe Bags may have a compromised or incomplete sterile pouch seal. Investigation determined that during the supplier manufacturing and packaging process, the Vision Probe Bag is placed into a separate sterile pouch for sealing. During this step, air trapped inside the Vision Probe Bag can remain within the pouch and expand during sterilization, which may place stress on the sterile-pouch seals and result in partial seal failure, creating a potentia
Quantity110580
DistributionU.S.: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MA, MD, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. O.U.S.: Australia, China, France, Germany, Italy, Japan, South Korea, Spain, Switzerl
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212048Ion Endoluminal System (Ion Fully Articulating Catheter), Ion Endoluminal System (Ion Peripheral Vision Probe)
Intuitive Surgical, Inc., Aug 26, 2021