Recall Z-1136-2026 ยท LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use

Sysmex America, Inc.

TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17

DateJan 21, 2026 (initiated Dec 15, 2025)
StatusOpen, Classified
ReasonUnder specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
Root causeSoftware design
ActionA Product Notification Field Action was sent to customers on 12/15/25. Actions for customers 1. To prevent injury, never place your hands around the lift section at any time. 2. In addition, please follow these instructions when handling rack jams or similar errors: " Pay special attention not to insert your hands around the lift section during any intervention. " If a rack, tube, or any object becomes lodged around the lift section, turn off the power. After confirming that the lift section has moved to the bottom dead point, carefully remove the rack and blood collection tubes. " If the rac
Quantity44 units
DistributionUS Nationwide.
FDA recordOpen on fda.gov