Recall Z-1340-2026 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan is a computer-based system designed for Leksell Gamma Knife treatment planning.

DateFeb 10, 2026 (initiated Dec 12, 2025)
StatusOpen, Classified
ReasonDue to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
Root causeUse error
ActionOn 12/12/2025, the firm sent an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" letter via email to customers informing them that when Defining a new stereotactic reference in an examination for a frame treatment while failing to first change an already defined obsolete stereotactic reference (representing a non-current frame fitting) to a pre-plan reference, may result in a treatment plan that irradiates the wrong patient location if the resulting inconsistent alignment between the stereotactic references is not detected by the user. Customers are instructed to: If th
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, Chin
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173791Leksell GammaPlan
Elekta Instrument AB, Feb 9, 2018