Recall Z-1087-2026 ยท LRO General Surgery Tray

Medline Industries, LP

Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9)

DateJan 14, 2026 (initiated Dec 16, 2025)
StatusOpen, Classified
ReasonMedline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Root causeUnder Investigation by firm
ActionMedline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/16/2025 via email and USPS first class mail. The notice explained the issue, potential risk, and requested the following: "1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantityof affected product you have in inventory on the form. Even if you do not have any affected productin inventory, please
Quantity4536 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.