Recall Z-1086-2026 ยท FDE Laparoscopy Kit
Medline Industries, LP
Medline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CDS, REF CDS983678K; 4) LAPAROSCOPY PELV CDS, REF CDS983679K; 5) LAP CHOLE CDS-LF, REF CDS983752M; 6) LAP CHOLE CDS-LF, REF CDS984016B; 7) LAP CHOLE KIT, REF CDS984020P; 8) LAP CHOLE-LF, REF CDS984175J; 9) LAP COLON CDS, REF CDS984849N; 10)
| Date | Jan 14, 2026 (initiated Dec 16, 2025) |
| Status | Open, Classified |
| Reason | Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/16/2025 via email and USPS first class mail. The notice explained the issue, potential risk, and requested the following: "1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantityof affected product you have in inventory on the form. Even if you do not have any affected productin inventory, please |
| Quantity | 41432 units |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.