Recall Z-1069-2026 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Allura Xper FD10/10; Model Number: 722011;

DateJan 13, 2026 (initiated Dec 15, 2025)
StatusOpen, Classified
ReasonPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Root causeOther
ActionOn December 15, 2025, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU). b. Circulate this letter to all users so that they are aware of the issue. Keep this letter with the documentation of the system until Philips inspects and, where needed, corrects your system. c. Establish an emergency protocol prior to all applicable diagnostics, interventional and minimally invasive procedures to manage the situation should you experience the issue during a procedure. d.
Quantity21 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Ken
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102005ALLURA XPER OR TABLE SERIES
Philips Medical Systems North America, Inc., Aug 9, 2010
K130638ALLURACLARITY XPER FD SERIES X-RAY SYSTEM
Philips Medical Systems Nederland B.V., Jun 28, 2013
K130842ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES
Philips Medical Systems Nederland B.V., May 21, 2013
K133292ALLURA XPER FD; OR TABLE
Philips Medical Systems Nederland B.V., Mar 5, 2014
K141979ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Philips Medical Systems Netherland BV, Aug 19, 2014
K161563Allura Xper FD series and Allura Xper OR Table series
Philips Medical Systems Nederland B.V., Jul 29, 2016