Recall Z-1231-2026 ยท MDP Separator For Therapeutic Purposes, Membrane Automated Blood Cell/Plasma

VANTIVE US HEALTHCARE LLC

PRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System

DateFeb 2, 2026 (initiated Jan 6, 2026)
StatusOpen, Classified
ReasonThere is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Root causeComponent design/selection
ActionVantive issued an Urgent Medical Device Correction notice to its consignees on 01/06/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: 1. You may continue to perform therapy using the Prismaflex sets listed above. Please monitor the deaeration chamber during therapy to ensure it remains in an upright position within the holder. 2. Vantive is aware of customers who have experienced this issue and attempted to further secure the deaeration chamber in an upright position. If attempting to secure the deaeration chamber in an upright
Quantity13,760 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.