ClearanceBrief
Recall Z-1314-2026 ยท KNW Instrument, Biopsy

Hologic, Inc

Brevera Breast Biopsy System Disposable 9 Gauge Needle. Biopsy Instrument.

DateFeb 6, 2026 (initiated Dec 15, 2025)
StatusOpen, Classified
ReasonPresence of particulates in affected devices that can be deposited into breast tissue during use.
Root causeUnder Investigation by firm
ActionAn URGENT FIELD SAFETY NOTICE (REMOVAL) was mailed to consignees on 1/2/26. The notification instructs consignees to quarantine affected devices and return them through IQVIA per the instructions provided on the Recall Response Form. The recall notification should be forwarded within consignee locations to users and to any facility to which product was further distributed. A copy of the recall notification is to be posted for awareness at consignee locations. Questions can be directed to BreastHealth.Support@hologic.com or by phone at 1-877-371-4372. An updated FIELD TECHNICAL BULLETIN dated
Quantity681,870 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163052Brevera Biopsy System with Corlumina Imaging Technology
Hologic, Inc., Mar 29, 2017