ClearanceBrief
Recall Z-1129-2026 ยท OEQ Angiography/Angioplasty Kit

Merit Medical Systems, Inc.

Allwell Angioplasty Pack REF: IS-30-B1/B

DateJan 16, 2026 (initiated Dec 4, 2025)
StatusOpen, Classified
ReasonInflation device handle may detach from the syringe during procedure.
Root causeProcess control
ActionOn December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the produc
Quantity328000 units
DistributionWorldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.