Recall Z-1127-2026 ยท OEZ Cardiovascular Procedure Kit
Merit Medical Systems, Inc.
Merit Medical , Custom Manifold Kit REF: K09-13203A
| Date | Jan 16, 2026 (initiated Dec 4, 2025) |
| Status | Open, Classified |
| Reason | Inflation device handle may detach from the syringe during procedure. |
| Root cause | Process control |
| Action | On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the produc |
| Quantity | 2192 units |
| Distribution | Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.